Test Frequencies 100Hz, 120 Hz, 1 kHz, 10 kHz, 100 kHz, 200 kHz
Frequency Accuracy ±0.1%
Test Levels 1 Vrms, 0.25 Vrms, 0.05 Vrms, 1 Vdc (for DCR)
Level Accuracy ±10%
Output Impedance 100 , ±5%
Measurement Range
Impedance (Z)
Frequency Maximum Minimum Best Resolution
DCR 20 M 0.1 0.001
100 Hz 20 M 0.1 0.001
120 Hz 20 M 0.1 0.001
1 kHz 20 M 0.1 0.001
10 kHz 20 M 0.1 0.001
100 kHz 10 M 0.1 0.001
200 kHz 10 M 0.1 0.001
Capacitance (C)
Frequency Maximum Minimum Best Resolution
100 Hz 15.91 mF 79.57 pF 0.001
120 Hz 13.26 mF 66.31 pF 0.001
1 kHz 1.591 mF 7.957 pF 0.001
10 kHz 159.1 µF 0.795 pF 0.001
100 kHz 15.91 mF 0.159 pF 0.001
200 kHz 7.957 µF 0.079 pF 0.001
Inductance (L)
Frequency Maximum Minimum Best Resolution
100 Hz 31.83 kH 159.1 µH 0.001
120 Hz 26.52 kH 132.6 µH 0.001
1 kHz 3.183 kH 15.91 µH 0.001
10 kHz 318.3 H 1.591 µH 0.001
100 kHz 31.83 H 0.159 µH 0.001
200 kHz 15.91 H 0.079 µH 0.001
GENERAL
Operating Temperature 32˚ to 104˚F (0˚ to 40˚C)
Storage Temperature -4˚ to 158˚F (-20˚ to 70˚C)
Relative Humidity up to 85%
AC Operation 110V/220V AC, 60/50Hz
Dimensions (LxWxH) 11.8 x 8.7 x 5.9" (300 x 220 x 150mm)
Weight 10 lbs (4.5 kg)
Supplied Accessories: User manual, Kelvin probe, DMM test leads, and power cord
Specifications
889B
Three Year Warranty
model
The B&K Precision 889B synthesized in-circuit bench LCR/ESR meter is a
high accuracy test instrument used for measuring inductors, capacitors and
resistors with a basic accuracy of 0.1%. Also, with the built-in functions of
DC/AC Voltage/Current measurements and Diode/Audible Continuity checks,
the 889B can not only help engineers and students to understand the
characteristics of electronics components, but also being an essential tool on
any service bench.
Features:
Measurement parameters: Z, L, C, DCR, ESR, D, Q, ACV, DCV, ACA,
DCA and Ø
Test conditions: 100 Hz, 120 Hz, 1 kHz, 10 kHz, 100 kHz, 200 kHz,
1 Vrms, 0.25 Vrms, 0.05 Vrms
0.1% basic accuracy
Dual LCD display
Very quick response, user-friendly
Fully auto/manual selection
DC resistance measurement
USB interface