Chapter 7 Specifications
A-1 Introduction
Chapter 7 describes the complete specifications of the multimeter.
A-2 Technical Specifications
Specifications are based on the following conditions:
One year calibration cycle.
Operating temperature at 18 °C to 28 °C
Accuracy is expressed as: ± (% of reading + % of range) after 30 minutes warm-up.
Temperature Coefficient: add ± [0.1% × (the applicable accuracy)/°C] for 0 °C to 18 °C and
28 °C to 40 °C
Relative Humidity: Up to 80% RH for 0 °C to 28 °C (75% RH for 10 MΩ and above ranges
for resistance measurement). Up to 70% RH for 28 °C to 40 °C
Displayed Readings and Reading rates
Reading rates on front panel (approximate reading/sec)
Measurement Function
Slow
Med
Fast
DCV
5
10
25
DCA
5
10
25
ACV
5
10
25
ACA
5
10
25
Ω (Ranges below 2M )
5
10
25
Ω (20M range and above)
1.3
2.6
5.6
Freq/ Period
1
2
3.9
Using dual display
0.9
0.9
0.8
True RMS DC+AC
1.2
1.4
1.5
Diode
————
10
————
Continuity
————
————
25